Metallurgical · hardness

Hardness tester calibration (Brinell, Rockwell, Vickers)

Foundry and metallurgical labs live and die by hardness numbers. Keeping a hardness tester honest means direct verification of what it applies and indirect verification of what it reads.

By Versatile · equipment manufacturer & calibration provider · updated 2026-07-15

The governing standards

ScaleISOASTM
Brinell (HB)ISO 6506ASTM E10
Rockwell (HR)ISO 6508ASTM E18
Vickers (HV)ISO 6507ASTM E92 / E384

Each standard has multiple parts - the machine, the test blocks and the indenters are covered separately. Calibration and verification follow those parts.

Direct verification

Direct verification checks the machine's own elements against traceable references: the test force at each level, the indenter geometry, and the measuring system (depth for Rockwell, optical for Brinell/Vickers). It is typically done at installation, after a move or repair, and periodically.

Indirect verification with test blocks

For a tester in daily use, indirect verification with certified reference test blocks is the everyday control. A common procedure (per ISO 6508-2 / ASTM E18) is:

Test blocks are directional - test on the certified face only, near (not on) previous indents, and never re-use an indent site. A scarred block gives false failures.

How often

EnvironmentTypical calibration frequency
High-usage production lab3-6 months
Quality-control laboratory6-12 months
Reference laboratory≤ 6 months
R&D / low usage12 months

A daily indirect check on one block per scale in use is good practice regardless of the formal interval.

Common problems

Frequently asked questions

What standards govern hardness tester calibration?
ISO 6506 (Brinell), ISO 6508 (Rockwell) and ISO 6507 (Vickers), with the ASTM equivalents E10, E18 and E92/E384. Each has parts covering the machine, the test blocks and the indenters.
What is the difference between direct and indirect verification?
Direct verification checks the machine’s own elements - test force, indenter geometry and measuring system - against traceable references, usually at installation or after repair. Indirect verification uses certified reference test blocks (typically five readings on three blocks per scale, low/mid/high) and is the routine, in-service control.
How often should a hardness tester be calibrated?
Roughly every 3-6 months in a high-usage production lab, 6-12 months in a QC lab, 12 months for low usage, and at most 6 months for reference laboratories - supported by a daily indirect check on a certified block.